ELI 100TM:
System Specifications
| |
ELI 100TM
Electrocardiograph
|
| Feature |
Specifications |
| Instrument
Type |
12-Lead Electrocardiograph |
| Input
Channels |
Simultaneous acquisition of
all leads |
| Standard
Leads Acquired |
I,II,III,aVR, aVL,aVF,
V1,V2,V3,V4,V5,V6 |
| Input
Impedance |
Meet or
exceed the
requirements of
ANSI / AAMI EC11 |
| Input
Dynamic Range |
| Electrode
Offset Tolerance |
| Frequency
Response |
| Patient
Leakage Current |
Meet or
exceed the
requirements of
ANSI / AAMI ES-1 |
| Chassis
leakage current |
| A/D
Conversion |
20 bits (1.17 microvolt
LSB) |
| Digital
Sampling Rate |
10,000 s/sec/channel used
for pacemaker spike detection.
500 s/sec/channel used
for recording and analysis |
| Special
Functions |
Baseline filter, lead off
and/or arifact detection, AC interference
rejection |
| Recording
Technique |
Computer controlled thermal
dot array (200 dots/inch) |
| Paper Type |
Thermal sensitive, full
grid,
108mm x 30m |
| Thermal
Writer Speeds |
5, 10, 25, 50 mm/s |
| Sensitivity |
5, 10, or 20 mm/mV |
| Device
Classification |
Type I, Class CF |
| Weight |
11 pounds (4.9 kg) |
| Dimensions |
11.25" x 11.50" x
3.75"
(28.1 x 28.7 x 9.3cm) |
| Power
Requirements |
100-240VAC at 50 or 60Hz,
or internal batteries (more than 1,000 ECGs per
battery charge) |
|
|
|