ELI 250TM:
System Specifications
|
|
ELI 250TM
Electrocardiograph
|
| Feature |
Specifications |
| Instrument
Type |
12-Lead Electrocardiograph |
| Input
Channels |
Simultaneous acquisition of
12 leads |
| Standard
Leads Acquired |
I, II, III, aVR, aVL, aVF,
V1,V2,V3,V4,V5,V6 |
|
Waveform Display |
Backlit, 1/4 VGA LCD |
| Input
Impedance |
Meet or
exceed the
requirements of
ANSI / AAMI EC11 |
| Input
Dynamic Range |
| Electrode
Offset Tolerance |
|
Common Mode
Rejection |
| Patient
Leakage Current |
Meet or
exceed the
requirements of
ANSI / AAMI ES 1 |
| Chassis
leakage current |
| Digital
Sampling Rate |
10,000 s/sec/channel used
for pacemaker spike detection.
1,000 s/sec/channel used
for recording and analysis |
| Special
Functions |
Optional Mortara VERITAS resting interpretation algorithm;
connectivity options for bi-directional communication; flash
memory storage of ECGs |
|
Thermal Recorder |
Computer-controlled
dot array; 8 dots/mm |
| Thermal
Writer Speeds |
5, 10, 25, 50 mm/s |
|
Gain Settings |
5, 10, or 20 mm/mV |
|
Report Print Formats |
3+1, 3+3 channel,
6-channel, 12-channel |
|
Rhythm Print Formats |
6-channel with
configurable lead groups or 12-channel |
|
Keyboard Type |
Elastomer keypad
with complete alphanumeric keys, soft-key menu and dedicated
function keys |
|
Frequency Response |
.05 to 300 Hz |
|
Filters |
High performance
baseline filter; AC interference filter 50/60 Hz; low pass
filters 40 Hz or 150 Hz |
|
A/D Conversion |
20 bit (1.17
microvolt LSB) |
| Device
Classification |
Class I, Type CF defibrillation proof applied parts |
|
ECG Storage |
Up to 60 ECGs |
| Weight |
11.25 lb. (5.1 kg)
|
| Dimensions |
15.5" x 17" x 4"
(39.4 cm x 43.2 cm x 10.2 cm) |
| Power
Requirements |
Universal AC power supply (100-240 VAC at 50/60 Hz) 50 VA.. Internally rechargeable battery |
|
|
|