Landscape
System Specifications
| |
LandscapeTM
Electrocardiograph
|
| Feature |
Specifications |
| Instrument
Type |
12-lead Electrocardiograph |
| Input
Channels |
Simultaneous acquisition of
all leads |
| Standard
Leads Acquired |
I,II,III,aVR, aVL,aVF,
V1,V2,V3,V4,V5,V6 |
| Display
type |
10.4" LCD |
| Rhythm
Display |
Up to full 12 leads |
| Input
Impedance |
Meet or
exceed the
requirements of
ANSI / AAMI EC11 |
| Input
Dynamic Range |
| Electrode
Offset Tolerance |
| Common Mode
Rejection |
| Frequency
Response |
| Patient
Leakage Current |
Meet or
exceed the
requirements of
ANSI / AAMI ES-1 |
| Chassis
Leakage Current |
| Digital
Sampling Rate |
10,000 s/sec/channel used
for pacemaker spike detection500 s/sec/channel
used for recording and analysis |
| Device Type |
Type I, Class CF |
| Special
Functions |
Baseline filter; AC
interference rejection; source consistency
filter; flash memory storage;
serial comparison |
| Recording
Technique |
Computer-controlled thermal
dot array; 8 dots/mm |
| Paper Type |
Thermal sensitive; grid;
perforated, Z-fold, or plain paper |
| Thermal
Writer Speeds |
5, 10, 25, 50 mm/s |
| Sensitivity |
5, 10, or 20 mm/mV |
| Weights |
Landscape acquisition
module:
6 lb. (2.72 kg) (includes battery)
With thermal writer:
10 lb. (4.54kg) |
| Dimensions |
Acquisition module:
10.9"" L x 11.4"" D x
1.5"" H;(27.7cm x 28.9cm x 3.8cm)
Acquisition module with thermal writer:
13.7"" L x 11.4"" D x
3.0"" H(34.8cm x 28.9cm x 7.6cm) |
| Power
Requirements |
Removable, non-spillable
lead-acid, 12 V; 100-240 V AC at 50 / 60 Hz |
|
|
|